用于评估植入式脉冲发生器的起搏器系统分析仪的局限性。

Limitations of pacemaker system analyzers for the evaluation of implantable pulse generators.

作者信息

Hauser R G, Edwards L M, Giuffre V W

出版信息

Pacing Clin Electrophysiol. 1981 Nov;4(6):650-7. doi: 10.1111/j.1540-8159.1981.tb06248.x.

Abstract

A variety of commercial pacemaker system analyzers have been used to evaluate the electrical performance of pulse generators prior to implant or at the time of reoperation for pacemaker malfunction. We evaluated the ability of seven commercially available pacemaker system analyzers to provide the data required to determine if a given pulse generator was operating according to its manufacturer's specifications. While all pacemaker system analyzers provided accurate data for rate and pulse duration, values displayed for amplitude and sensitivity were inconsistent due to the methods each pacemaker system analyzer employed to measure these parameters. Overall, 22% of representative test sequences yielded data which would have resulted in the inappropriate rejection of the subject pulse generators for implantation. These observations suggest that pacemaker system analyzers are not suitable for evaluating all pulse generators. Moreover, physicians must be aware of these deficiencies in order to avoid the pitfalls of relying exclusively on pacemaker system analyzer data when troubleshooting pacemaker malfunction.

摘要

在植入起搏器之前或因起搏器故障进行再次手术时,已使用多种商用起搏器系统分析仪来评估脉冲发生器的电气性能。我们评估了七种商用起搏器系统分析仪提供所需数据的能力,以确定给定的脉冲发生器是否按照其制造商的规格运行。虽然所有起搏器系统分析仪都能提供准确的心率和脉冲持续时间数据,但由于每种起搏器系统分析仪测量这些参数所采用的方法,显示的幅度和灵敏度值并不一致。总体而言,22%的代表性测试序列产生的数据会导致对受试脉冲发生器植入的不当拒绝。这些观察结果表明,起搏器系统分析仪并不适用于评估所有脉冲发生器。此外,医生必须意识到这些缺陷,以避免在排查起搏器故障时仅依赖起搏器系统分析仪数据所带来的陷阱。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验