Corrosion of pacemaker electrodes.

作者信息

Parsonnet V, Villanueva A, Driller J, Bernstein A D

出版信息

Pacing Clin Electrophysiol. 1981 May;4(3):289-96. doi: 10.1111/j.1540-8159.1981.tb03697.x.

Abstract

As improvements in pacemaker components and design increase pacemaker life, problems with other components may emerge such as corrosion of the electrodes. Explanted electrodes were examined under a scanning electron microscope and the degree of corrosion was graded and correlated with the reasons for explantation, the status of the explanted pacemaker, and the duration of the implant. Fifty-six explanted electrodes were removed for loss of capture or sensing, pacemaker extrusion, broken wires, or after death. A numerical grading system was used to qualitate the degree of corrosion. The pacemaker itself was retrieved in 50 cases, permitting evaluation of the entire pacing system. The dc component of the pacemaker output was measured. Without the aid of magnification most electrodes appeared shiny and smooth. With one exception all corrosion was microscopic. Significant corrosion was seen on all pacemaker electrodes that had been connected to pacemakers with dc offset currents greater than 5 microamperes. All except the "youngest" explanted electrode showed some degree of corrosion. The data suggest that corrosion was directly related to the duration of implantation. With the possible exception of one case, no consistent adverse clinical effects were observed, even when electrodes were heavily corroded. Whether or not corrosion eventually will progress to the point that clinical problems become manifest cannot be predicted at this time.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验