Barlow J S, Rémond A
Electroencephalogr Clin Neurophysiol. 1981 Nov;52(5):418-23. doi: 10.1016/0013-4694(81)90024-9.
A multichannel nulling or subtraction technique for minimizing or eliminating vertical and/or horizontal eye movement artifact including blinks in EEG recordings is described. The circuit, which makes use of principles previously described by other workers, is intended to function integrally with contemporary electroencephalographs, and is relative simple in design and easy to implement.
描述了一种用于最小化或消除脑电图记录中的垂直和/或水平眼动伪迹(包括眨眼)的多通道归零或减法技术。该电路利用了其他研究人员先前描述的原理,旨在与当代脑电图仪整体配合工作,并且设计相对简单且易于实现。