Nathan A, Hellestrand K, Bexton R, Nappholz T, Spurrell R, Camm J
Pacing Clin Electrophysiol. 1982 Mar;5(2):201-7. doi: 10.1111/j.1540-8159.1982.tb02215.x.
An external version of a new pacemaker designed for automatic tachycardia termination is described. In response to a tachycardia, defined as four successive beats occurring at a rate faster than a variable preset value, a number of stimuli (1-15) are generated. The initial coupling interval and subsequent pacing cycle intervals (where applicable) are always the same. Failure to terminate tachycardia results in the reduction of the pacing cycle length by 6 ms before termination of tachycardia is re-attempted. Pacing cycle length reduction may be repeated up to 16 times. If all coupling intervals of a single extrastimulus fail to achieve tachycardia termination, a second stimulus is scanned with the first through the same range of decrements. Additional stimuli, to a total of 15, could be used. A memory function is incorporated to retain and re-use a successful sequence. The pacemaker was evaluated in 16 patients with recurrent "supraventricular" (atrioventricular and intra A-V nodal) tachycardia, using right ventricular pacing. Termination was successful in all patients. More than two stimuli were necessary in only two patients, one of whom required five, and the other six stimuli. No unwanted arrhythmias were induced.
本文介绍了一种专为自动终止心动过速而设计的新型起搏器的体外模式。当检测到心动过速(定义为连续四个心跳的速率超过可变预设值)时,会产生多个刺激脉冲(1 - 15个)。初始耦合间期及后续起搏周期间期(如适用)始终保持一致。若未能终止心动过速,则在再次尝试终止心动过速前,将起搏周期长度缩短6毫秒。起搏周期长度的缩短最多可重复16次。若单个额外刺激的所有耦合间期均未能终止心动过速,则会扫描第二个刺激,并使其与第一个刺激以相同的递减范围进行扫描。总共可使用15个额外刺激。起搏器具备记忆功能,可保留并重复使用成功的刺激序列。使用右心室起搏对16例复发性“室上性”(房室及房室结内)心动过速患者的该起搏器进行了评估。所有患者的心动过速均成功终止。仅2例患者需要超过两个刺激脉冲,其中1例需要5个,另1例需要6个刺激脉冲。未诱发任何不良心律失常。