Fyhrquist F, Puutula L
Clin Chem. 1978 Jan;24(1):115-8.
We applied a 1-h radioimmunoassay incubation at 37 degrees C in determining generated angiotensin I in an assay for plasma renin activity. Under these nonequilibrium conditions, 26% of the 125I-labeled angiotensin I was bound at zero dose of unlabeled angiotensin, as compared to 57% at equilibrium after 18 h at 4 degrees C. Sensitivity and useful range for the standard curve remained unchanged. Blanks were not altered. There was a good (r = .971) correlation between renin values in 120 plasma samples from hypertensive patients as measured with both procedures. With isoelectric focusing, we detected no damage to the labeled angiotensin I during incubation for 1 h at 37 degrees C in the presence of diluted plasma, disodium ethylenediaminetetraacetate, hydroxyquinoline, and neomycin. Analytical recovery of unlabeled angiotensin I added to the assay mixture was 98 +/- 2.3% (mean +/- SD). We conclude that our incubation conditions allow rapid and accurate assay of plasma renin activity to be completed in one working day.
在测定血浆肾素活性的试验中,我们在37℃下进行了1小时的放射免疫分析温育,以测定生成的血管紧张素I。在这些非平衡条件下,未标记血管紧张素零剂量时,125I标记的血管紧张素I有26%被结合,相比之下,4℃下18小时达到平衡时为57%。标准曲线的灵敏度和有效范围保持不变。空白值未改变。两种方法测定的120份高血压患者血浆样本中的肾素值之间存在良好的相关性(r = 0.971)。通过等电聚焦,我们发现在37℃下,在存在稀释血浆、乙二胺四乙酸二钠、羟基喹啉和新霉素的情况下温育1小时,标记的血管紧张素I未受到损伤。添加到分析混合物中的未标记血管紧张素I的分析回收率为98±2.3%(平均值±标准差)。我们得出结论,我们的温育条件能够在一个工作日内快速准确地完成血浆肾素活性的测定。