分体探测器计算机断层扫描:初步报告。

Split-detector computed tomography: a preliminary report.

作者信息

Brooks R A, Di Chiro G

出版信息

Radiology. 1978 Jan;126(1):255-7. doi: 10.1148/126.1.255.

Abstract

A split-detector system, consisting of a thin CaF2 scintillator followed by a thick Nal crystal, was installed in an EMI Mark I scanner. Energy discrimination is possible because the CaF2 crystal responds primarily to low-energy photons; dual-energy images can be obtained from only one scan. Calculations and preliminary experiments show that the split-detector separates energies better than the 100-140 kVp method. It was possibel with the split detector to differentiate between two weak solutions of CaCl2 and Kl which had the same computed tomography number (44 H), under normal scanning conditions.

摘要

一种由薄的CaF2闪烁体和厚的碘化钠晶体组成的分裂探测器系统安装在EMI Mark I扫描仪中。由于CaF2晶体主要对低能光子作出反应,因此可以进行能量鉴别;仅通过一次扫描就能获得双能图像。计算和初步实验表明,分裂探测器比100 - 140 kVp方法能更好地分离能量。在正常扫描条件下,使用分裂探测器能够区分两种具有相同计算机断层扫描值(44 H)的氯化钙和碘化钾稀溶液。