False inhibition of an atrial demand pacemaker caused by an insulation defect in a polyurethane lead.

作者信息

van Gelder L M, El Gamal M I

出版信息

Pacing Clin Electrophysiol. 1983 Sep;6(5 Pt 1):834-9. doi: 10.1111/j.1540-8159.1983.tb04403.x.

Abstract

A patient who had an atrial demand pacemaker (AAI) presented with irregular pacing at a routine examination 5 months after implantation. When a magnet was applied over the pulse generator regular fixed rate pacing was obtained, thus proving oversensing in the system. Reprogramming the input sensitivity level to 2.5 and 5.0 mV did not solve the problem. Programming the pulse generator to the triggered mode (AAT) showed acceleration of the stimulation rate but also inhibition of the system. An S-S interval of 1260 ms was measured at a programmed interval of 857 ms (70 bpm). The pulse generator was disconnected and the intra-atrial electrogram was recorded. This showed different spurious signals varying in morphology and amplitude. Fortunately we were able to remove the lead (Medtronic 6991-U) from the atrial appendage. Subsequently a Helifix 12 mm AT lead was successfully implanted in the right atrial appendage and the same pulse generator was connected to the newly implanted lead. When the removed lead was examined by the manufacturer, a small tear in the insulation of the wire was detected. The dimensions of the tear were 0.1 X 0.7 mm. The tear was caused by stress corrosion cracking in the polyurethane tubing of the lead.

摘要

相似文献

1
False inhibition of an atrial demand pacemaker caused by an insulation defect in a polyurethane lead.
Pacing Clin Electrophysiol. 1983 Sep;6(5 Pt 1):834-9. doi: 10.1111/j.1540-8159.1983.tb04403.x.
2
Self-inhibition of an AV sequential demand (DVI) pulse generator due to polyurethane lead insulation disruption.
Pacing Clin Electrophysiol. 1983 Sep;6(5 Pt 1):840-4. doi: 10.1111/j.1540-8159.1983.tb04404.x.
3
Galvanic element produced by defective electrode insulation--a possible cause of abnormal pulse generator inhibition.
Scand J Thorac Cardiovasc Surg. 1983;17(1):85-8. doi: 10.3109/14017438309102385.
4
P wave oversensing in a unipolar VVI pacemaker.
Pacing Clin Electrophysiol. 1995 Feb;18(2):370-3. doi: 10.1111/j.1540-8159.1995.tb02529.x.
5
Suppression of ventricular output by noise in the atrial channel in a dual chamber pacemaker.
Pacing Clin Electrophysiol. 1995 Aug;18(8):1586-8. doi: 10.1111/j.1540-8159.1995.tb06746.x.
6
Chronic steroid-eluting lead performance: a comparison of atrial and ventricular pacing.
Pacing Clin Electrophysiol. 1997 Jan;20(1 Pt 1):17-24. doi: 10.1111/j.1540-8159.1997.tb04807.x.
7
VDD pacing with a previously implanted single lead system.
Pacing Clin Electrophysiol. 1992 Oct;15(10 Pt 1):1425-9. doi: 10.1111/j.1540-8159.1992.tb02914.x.
9
Autointerference of demand pulse generators.
Pacing Clin Electrophysiol. 1981 May;4(3):274-80. doi: 10.1111/j.1540-8159.1981.tb03695.x.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验