Suppr超能文献

Galvanic element produced by defective electrode insulation--a possible cause of abnormal pulse generator inhibition.

作者信息

Levander-Lindgren M, Sylvén J C, Thorén A

出版信息

Scand J Thorac Cardiovasc Surg. 1983;17(1):85-8. doi: 10.3109/14017438309102385.

Abstract

In patients with pacemaker, abnormal inhibition with prolongation of pacing intervals may cause alarming clinical symptoms. A case is described in which high current threshold in relation to voltage threshold indicated probability of an insulation defect with current leakage. Electrograms from the electrode disclosed false signals, which had appeared after replacement of a pulse generator six months earlier. A sharp bend of the wire in the tricuspid area was shown by X-ray and was accentuated by movements of the valve. Experimentally it was demonstrated that similar potentials, sufficient to inhibit a pulse generator, can be obtained from an electrode with defective insulation. A galvanic element forms between the metals of the electrode tip and the non-insulated cable, and potential variations are elicited by movement of the wire.