J Hosp Infect. 1984 Jun;5(2):155-63. doi: 10.1016/0195-6701(84)90119-1.
The growth of Escherichia coli over a period of 8 h under simulated ward conditions was compared in feeding systems containing either Clinifeed ISO or Nutrient Broth. In both, counts increased from 10(2) to 10(7) ml-1. Other systems containing Clinifeed ISO were inoculated with E. coli and sampled over 24 h. At 8 and 16 h the contaminated reservoir was replaced, refilled or replaced together with the gastric drip line. When the reservoir was replaced or refilled it was always recontaminated by residual organisms. Even when the reservoir and drip line were replaced, although the contents of the reservoir were sterile, E. coli was still detected in feed collected from the end of the fine-bore tube. Experiments with varying numbers of E. coli in the inoculum demonstrated that even one organism in the reservoir could, within 16 h, multiply to a level that might be harmful, especially to compromised patients.
在模拟病房条件下,对含有Clinifeed ISO或营养肉汤的喂养系统中大肠杆菌在8小时内的生长情况进行了比较。在这两种系统中,细菌数量均从10²增至10⁷ ml⁻¹。其他含有Clinifeed ISO的系统接种大肠杆菌后进行了24小时的采样。在8小时和16小时时,更换、重新填充或与胃滴注管一起更换受污染的储液器。当储液器被更换或重新填充时,它总是会被残留的微生物重新污染。即使储液器和滴注管都被更换,尽管储液器中的内容物是无菌的,但在从细孔管末端收集的饲料中仍检测到大肠杆菌。对接种物中不同数量大肠杆菌进行的实验表明,即使储液器中有一个细菌,在16小时内也可能繁殖到有害水平,尤其是对身体虚弱的患者。