Waller B F, Girod D A, Dillon J C
J Am Coll Cardiol. 1984 Dec;4(6):1235-41. doi: 10.1016/s0735-1097(84)80143-6.
Aortic wall tears resulting from rupture of dilating balloons in infants undergoing transluminal balloon angioplasty have not been previously recognized. A 2 day old infant had percutaneous transluminal balloon angioplasty to dilate a stenotic aortic valve, and after multiple balloon inflations the balloon burst. The infant died 2 days after undergoing transluminal balloon angioplasty and at necropsy a circumferential, transverse aortic wall tear was found. To test the hypothesis that rupture of appropriately sized balloons results in similar aortic wall tears, six unfixed, intact infant aortas were subjected to transluminal balloon angioplasty at necropsy: two infants had balloon rupture with inflated balloon diameter similar to that of the aorta, two had balloon rupture with an undersized balloon and two had dilation with an oversized balloon. Transverse wall tears occurred in aortas with similar aortic and balloon diameters; no aortic wall damage occurred with rupture of undersized balloons, and aortic rupture resulted from the use of oversized balloons. Thus, intimal-medial tears in the infant aorta may result from balloon bursting during angioplasty when aortic and inflated balloon diameters are similar.
在接受经腔球囊血管成形术的婴儿中,扩张球囊破裂导致的主动脉壁撕裂此前尚未被认识到。一名2日龄婴儿接受经皮经腔球囊血管成形术以扩张狭窄的主动脉瓣,在多次球囊充气后球囊破裂。该婴儿在接受经腔球囊血管成形术后2天死亡,尸检时发现主动脉壁有环形横向撕裂。为了验证适当尺寸的球囊破裂会导致类似主动脉壁撕裂的假设,在尸检时对六个未固定的完整婴儿主动脉进行经腔球囊血管成形术:两名婴儿的球囊破裂时充气球囊直径与主动脉相似,两名婴儿使用尺寸过小的球囊破裂,两名婴儿使用尺寸过大的球囊进行扩张。在主动脉和球囊直径相似的主动脉中出现横向壁撕裂;尺寸过小的球囊破裂未造成主动脉壁损伤,而使用尺寸过大的球囊导致主动脉破裂。因此,当主动脉和充气球囊直径相似时,婴儿主动脉的内膜-中层撕裂可能是血管成形术期间球囊破裂所致。