Suppr超能文献

使用带有数据采集系统的曲线再生功能的火焰III型自动分析仪和火焰IV型自动分析仪对血清电解质结果进行的比较。

A comparison of serum electrolyte results by the Flame III Auto-Analyser and a Flame IV Auto-Analyser which utilises curve regeneration with a data acquisition system.

作者信息

MacAulay M A, Mathers J J, Jacklyn C J

出版信息

Clin Biochem. 1978 Feb;11(1):23-7. doi: 10.1016/s0009-9120(78)80039-3.

Abstract

Results obtained with a Technicon Flame Photometer Model IV Auto-Analyzer, which utilizes curve regeneration in conjunction with a mini-computer for the continuous-flow determination of serum electrolytes, is presented. The system, featuring optimal flow conditions, subjects sample peaks to a mathematical treatment to effectively remove exponential deformation. The evaluation, undertaken after 2 months of operation in the routine clinical laboratory, utilizes the Flame Photometer Model III as the "referee" procedure. The comparison includes a statistical analysis of quality control reference material, simultaneously determined patient data and an evaluation of repeated patient data. The system gave comparable results to those obtained for the Flame Photometer Model III and provides the additional benefits of increased sample throughput, decreased reagent comsumption and sample interaction and gives reductions in the clerical tasks associated with the retrieval of data from Auto-Analyzer systems.

摘要

本文展示了使用Technicon IV型火焰光度计自动分析仪获得的结果,该分析仪利用曲线再生结合小型计算机对血清电解质进行连续流动测定。该系统具有最佳流动条件,对样品峰进行数学处理以有效消除指数变形。在常规临床实验室运行2个月后进行评估,使用III型火焰光度计作为“裁判”程序。比较包括对质量控制参考物质的统计分析、同时测定的患者数据以及对重复患者数据的评估。该系统给出的结果与III型火焰光度计相当,还具有增加样品通量、减少试剂消耗和样品相互作用的额外优点,并减少了与从自动分析仪系统检索数据相关的文书工作。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验