Jacklyn C L, MacAulay M A, Mathers J M, Munro C A, Kriz C D
Clin Biochem. 1978 Feb;11(1):16-22. doi: 10.1016/s0009-9120(78)80038-1.
Curve regeneration is applied to the continuous-flow determination of serum Na +, K +, Cl- and CO2 by the Flame photometer IV (Technicon Corp.). A Hewlett-Packard 2100 A mini-computer is used for data acquisition. The continuous-flow parameters of both rise and fall curves are estimated from computer sampled voltage outputs over the standard profile. An additional interaction correction variable designated as Beta (beta) is described and applied to the regenerated peaks of the cresol red CO2 procedure. The phenolphthlalein CO2 methodology, showing improved flow parameters over those for the cresol red procedure is adapted for the Flame IV system. The basic program design is briefly outlined. Typical computer determined calibration curves (linear regression) and sample peak tracings for both the basic and regenerated techniques are illustrated.