A conventional gas chromatograph with one system for split and splitless injection and one on-column injection system for fused-silica capillary columns and the two selective nitrogen-phosphorus and electron-capture detectors is applied to pesticide residue analysis in food samples. The gas chromatograph is equipped with a two-channel data processor that can be programmed with BASIC. Additionally an autosampler is used with the splitless injector. This automated injection system is connected to a 25-m methylsilicone fused-silica column which is coupled via an effluent splitter parallel to both detectors and used for screening in routine analysis. Calibration is performed on this column by means of three calibration test mixtures which include three internal standards. All compounds are calibrated on both detectors in parallel and the response calculated as an additional identification parameter. After the analysis of a of food samples together with the calibration mixture a report is plotted, containing all pesticide residues which may be present in the samples and their tentative quantities. The final confirmation is achieved on the second column, coated with methylphenyl-silicone phase, connected to the on-column injector.
一台常规气相色谱仪,配有一个用于分流和不分流进样的系统以及一个用于熔融石英毛细管柱的柱上进样系统,还有两个选择性氮磷检测器和电子捕获检测器,用于食品样品中的农药残留分析。该气相色谱仪配备了一个可使用BASIC编程的双通道数据处理器。此外,在不分流进样器上使用了自动进样器。这个自动进样系统连接到一根25米长的甲基硅氧烷熔融石英柱上,该柱通过一个流出物分流器与两个检测器并联,用于常规分析中的筛选。通过包含三种内标的三种校准测试混合物在该柱上进行校准。所有化合物在两个检测器上并行校准,并将响应计算为一个额外的鉴定参数。在分析了一批食品样品以及校准混合物后,绘制一份报告,其中包含样品中可能存在的所有农药残留及其暂定含量。最终确证在连接到柱上进样器的涂有甲基苯基硅氧烷固定相的第二根柱上进行。