Alm S, Jonson S, Karlsson H, Sundholm E G
J Chromatogr. 1983 Jan 7;254:179-86. doi: 10.1016/s0021-9673(01)88331-x.
A convenient and rapid gas chromatographic method for analysis of drugs of abuse has been developed. By installing two differently coated columns in a common split-splitless injector and connecting the column ends to a nitrogen-phosphorus detector and a flame ionization detector, considerable and accurate chromatographic information could be obtained from a single run. Quantitative analyses of amphetamine and some other drugs have been performed with the same instrumental set-up. Calibration graphs were evaluated and showed linearity in the concentration range chosen (0.5-10 mg/ml). By using an auto-sampler in combination with a BASIC program, both types of analyses can be performed routinely.
已开发出一种用于分析滥用药物的简便快速的气相色谱法。通过在一个普通的分流-不分流进样器中安装两根不同涂层的色谱柱,并将柱端连接到氮磷检测器和火焰离子化检测器上,单次运行就能获得大量准确的色谱信息。使用相同的仪器装置对苯丙胺和其他一些药物进行了定量分析。评估了校准曲线,结果表明在所选择的浓度范围(0.5 - 10毫克/毫升)内呈线性。通过将自动进样器与一个BASIC程序结合使用,两种类型的分析都可以常规进行。