Dimitriou P, Fretzayas A, Nicolaidou P, Gritzali F, Kasfiki A, Karpathios T
J Nucl Med. 1984 Jul;25(7):792-5.
We estimate the absorbed dose to the bladder walls, Db, due to [99mTc]pertechnetate direct radionuclide cystography, with due consideration to the dynamic nature of the examination. Our analytical method resulted in Db values expressed as functions of the bladder filling time. These values are compared with published data based on a static bladder model. It is shown that the Db value depends strongly on the protocol under which the examination is carried out. It is also shown that the normal saline flow rate and the time of the administration of the radioactive material can be adjusted so that the radiation burden may be greatly reduced.
我们估算了因[99mTc]高锝酸盐直接放射性核素膀胱造影术导致的膀胱壁吸收剂量Db,并充分考虑了检查的动态特性。我们的分析方法得出的Db值表示为膀胱充盈时间的函数。将这些值与基于静态膀胱模型的已发表数据进行比较。结果表明,Db值在很大程度上取决于进行检查所采用的方案。还表明,可以调整生理盐水流速和放射性物质的给药时间,从而大大降低辐射负担。