Image resolution of a microchannel plate x-ray image intensifier.

作者信息

Gould R G

出版信息

Med Phys. 1978 Jan-Feb;5(1):27-30. doi: 10.1118/1.594463.

Abstract

The resolution has been evaluated for an experimental x-ray image intensifier employing three microchannel plates (MCPs) as the photon absorber and electron multipliers. The line spread function (LSF) was measured and used for determination of the modulation transfer function (MTF). The MTF was found to be independent of the incident photon energy from 20 to 150 keV. An additional measurement using a lead-bar test pattern showed that the resolution exceeded 7 line pairs (lp)/mm. The factors influencing the resolution capabilities of the intensifier are discussed. The resolution is limited primarily by the 53-micron center-to-center separation of the channels of the MCPs.