一种用于计算机断层扫描的胶片剂量测定系统。

A film dosimetry system for use in computed tomography.

作者信息

Dixon R L, Ekstrand K E

出版信息

Radiology. 1978 Apr;127(1):255-8. doi: 10.1148/127.1.255.

Abstract

The authors describe a film dosimetry system for use in calculating the surface dose delivered by a CT scanner. Kodak XV-2 film is wrapped around a cylindrical water-filled phantom and the dose distribution is recorded. This system is easier to use than thermoluminescent dosimetry (TLD) and provides a detailed map of the dose distribution. Comparison with TLD measurements for a variety of CT scanners indicates that an accuracy of +/-15% can be achieved using this system. Dose distributions obtained with several scanners are shown.

摘要

作者描述了一种用于计算CT扫描仪所传递表面剂量的胶片剂量测定系统。柯达XV - 2胶片缠绕在一个圆柱形充水体模上,并记录剂量分布。该系统比热释光剂量测定法(TLD)更易于使用,并提供剂量分布的详细图谱。与多种CT扫描仪的TLD测量结果比较表明,使用该系统可实现±15%的准确度。文中展示了用几种扫描仪获得的剂量分布情况。