Mamalis N, Apple D J, Brady S E, Notz R G, Olson R J
J Am Intraocul Implant Soc. 1984 Spring;10(2):191-9. doi: 10.1016/s0146-2776(84)80108-1.
We analyzed 18 explanted 91Z anterior chamber lenses by light and scanning electron microscopy. Intermittent touch of the lens loops to the posterior corneal surface and the anterior chamber angle may have occurred. Erosion of the polypropylene loops into the anterior chamber angle recess and into the iris tissue was also observed. Fibrous tissue, uveal tissue, and inflammatory debris were noted on the loops, forming dense synechias at the points of contact with the angle recess. In some lenses the edges of the optics were sharp. Other significant manufacturing defects were rarely seen, and there was no evidence of degradation of polypropylene loops. The problems regarding surgical removal of this lens are discussed.
我们通过光学显微镜和扫描电子显微镜对18枚取出的91Z前房型人工晶状体进行了分析。人工晶状体袢可能间歇性地接触到角膜后表面和前房角。还观察到聚丙烯袢侵蚀到前房角隐窝和虹膜组织中。在袢上发现有纤维组织、葡萄膜组织和炎性碎屑,在与房角隐窝的接触点形成了致密的粘连。在一些人工晶状体中,光学部边缘锐利。很少见到其他明显的制造缺陷,也没有聚丙烯袢降解的证据。本文讨论了该人工晶状体手术取出的相关问题。