[人工晶状体的转归:扫描电子显微镜研究]

[Fate of artificial lenses: a scanning electron microscopy study].

作者信息

Colin J, Fardoun T, Bonissent J F, Sourdille P, Sinquin J

出版信息

J Fr Ophtalmol. 1984;7(11):721-6.

PMID:6533164
Abstract

Thirty one intraocular lenses removed from human eyes 2 months to 30 years after implantation were examined by scanning electron microscopy. Binkhorst 4 loop lenses (12) and 2 loop lenses (6), and Choyce lenses (6) were the most prevalent styles examined. Most of the lenses produced before 1978 were found to have manufacturing defects (mold disalignment, unfinished gate areas, etc.). More recent lenses showed an improvement in the quality of the surface. The lenticular, nonimpinging portion of the lenses made with polymethylmethacrylate did not demonstrate signs of degradation. However the holes in the edges of the lenses were rough and poorly finished with obvious irregularities. On the other hand, the lens foot of a Choyce lens revealed severe unusual abnormalities: the plastic surface changes may be consistent with degradation. Examination of explanted iris clip lenses showed degradation of the polypropylene loops in two cases: the loops were found to have fractured surfaces.

摘要

对31枚植入后2个月至30年从人眼中取出的人工晶状体进行了扫描电子显微镜检查。所检查的最常见类型是Binkhorst 4袢晶状体(12枚)和2袢晶状体(6枚)以及Choyce晶状体(6枚)。发现1978年以前生产的大多数晶状体存在制造缺陷(模具错位、浇口区域未加工等)。较新的晶状体表面质量有所改善。由聚甲基丙烯酸甲酯制成的晶状体的晶状体部分、非撞击部分未显示降解迹象。然而,晶状体边缘的孔粗糙且加工不佳,有明显的不规则之处。另一方面,一枚Choyce晶状体的晶状体袢显示出严重的异常:塑料表面变化可能与降解一致。对取出的虹膜夹型晶状体的检查显示,在两例中聚丙烯袢发生了降解:发现袢有断裂表面。

相似文献

8
Scanning electron microscopy of removed intraocular lenses.取出的人工晶状体的扫描电子显微镜检查。
J Am Intraocul Implant Soc. 1983 Spring;9(2):176-83. doi: 10.1016/s0146-2776(83)80034-2.
9
Management of anterior segment complications of intraocular lenses.人工晶状体前段并发症的处理
Ophthalmology. 1980 Sep;87(9):881-6. doi: 10.1016/s0161-6420(80)35152-x.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验