Pla C, Heese R, Pla M, Podgorsak E B
Med Phys. 1984 Jul-Aug;11(4):539-46. doi: 10.1118/1.595524.
A single-field rotational total skin electron irradiation technique has recently been developed at the McGill University for treatment of skin malignancies. The dose received by a given surface point during rotation in a uniform large electron field depends on the radius of rotation of the surface point, on the local radius of curvature of the contour in the vicinity of the point of interest, and on the shadows cast by limbs (arms upon trunk or head and neck, and legs upon each other). A method for calculating the surface dose distribution on a patient is presented accounting for the various parameters affecting the dose. A series of measurements were performed with polystyrene and a humanoid phantom, and an excellent agreement between measured and calculated dose distributions was obtained.