Suppr超能文献

The titanium/silicone rubber clip for female sterilization.

作者信息

Filshie G M, Casey D, Pogmore J R, Dutton A G, Symonds E M, Peake A B

出版信息

Br J Obstet Gynaecol. 1981 Jun;88(6):655-62. doi: 10.1111/j.1471-0528.1981.tb01226.x.

Abstract

A composite titanium/silicone rubber clip has been developed for female sterilization. The 12.7 mm long metal clip is lined with silicone rubber and is bent round the Fallopian tube by means of an applicator that can be used laparoscopically. As tubal necorsis occurs, the rubber expands to keep the lumen blocked. The tube eventually divides and the stumps heal closed. Since the Mark 3 version of the clip was introduced, over 5700 women in several countries have been sterilized. The first 540 British women who were followed-up for five to 17 months experienced one method failure (ectopic) and tow operator failures, to total failure rate of 5.5/1000. Clip-related complications were minimal and long-term sequelae were comparable to tubal ligation. The clip causes minimal destruction but still seems effective for tubal occlusion and it can accommodate thick tubes.