Studebaker G A, Cox R M, Wark D J
Audiology. 1978 Mar-Apr;17(2):173-86. doi: 10.3109/00206097809080044.
The validity of probe tube microphone measurements in providing data indicative of the magnitude of the change in subjects' threshold resulting from changes in earmold configuration was investigated. The relationship between these measures in the real ear canal and the changes observed in 2-cm3 and Zwislocki couplers under similar circumstances or earmold modification was also measured. Standard, vented and 'open' earmold conditions were utilized. Threshold and probe-tube measurements were made. Statistical evaluation revealed that these two techniques did not produce significantly different results except at 125--165 Hz, where noise masking may have been a factor influencing the threshold data. Neither coupler as used gave an accurate quantitative estimate of the in-use effects of vented earmolds or the open earmold configuration, but the Zwislocki coupler gave a better approximation than the 2-cm3 coupler.