Soldin S J, Hill J G
Clin Chem. 1978 May;24(5):747-50.
We describe a procedure for the rapid and specific measurement of creatinine, in which it is separated from other compounds in serum or urine by paired-ion chromatography and is quantified by measuring its absorbance at 200 nm. The procedure can be done on as little as 10 microliter of serum. Between-day precision studies for concentrations of 13 and 62 mg/liter yielded coefficients of variation of 6.9 and 2.2%, respectively. Analytical recovery of various amounts of creatinine added to plasma exceeded 95% in all cases. The proposed procedure was compared with the continuous-flow procedure by analyzing a series of serum and urine specimens by both methods. There was excellent agreement for urine specimens, but with serum the results by the present method were significantly (P less than 0.001) lower.
我们描述了一种快速、特异性测定肌酐的方法,该方法通过离子对色谱法将血清或尿液中的肌酐与其他化合物分离,并通过测量其在200nm处的吸光度进行定量。该方法仅需10微升血清即可完成。对浓度为13和62mg/升的日间精密度研究,变异系数分别为6.9%和2.2%。在所有情况下,添加到血浆中的各种量肌酐的分析回收率均超过95%。通过两种方法分析一系列血清和尿液标本,将所提出的方法与连续流动法进行比较。尿液标本的结果一致性很好,但对于血清,本方法的结果明显更低(P小于0.001)。