Abstract

In a group of 40 patients discharged from the hospital after mitral valve replacement with an Ionescu-Shiley pericardial xenograft from January, 1977, to December, 1980, seven instances of valve failure occurred. This unprecedented incidence of bioprosthetic dysfunction after mitral valve replacement with the Ionescu-Shiley pericardial xenograft led us to update the follow-up of our patients receiving this valve in the mitral position. The results of our survey showed, at 6 years postoperatively, an actuarial survival rate of 72%, an actuarial probability of being free from emboli of 62%, and an actuarial freedom from prosthetic failure of 60%. The explanted valves showed complete absence of the neoendothelial lining of the Dacron-covered frame and lesions resembling those observed in Ionescu-Shiley pericardial xenografts removed from our fatigue test system after a mean of 29 +/- 17 X 10(6) cycles. This similarity prompted a classification of tears occurring in the mitral Ionescu-Shiley pericardial xenograft, which is of extreme importance, since the clinical presentation and outcome of patients with a failing valve differs according to the type of rupture. The lesions of the cusp observed in clinical specimens were possibly related to the continuous trauma of the tissue against the bare Dacron cloth during closure of the valve. It is concluded that (1) actual durability and thrombogenicity of the Ionescu-Shiley pericardial xenograft in the mitral position needs to be carefully reassessed, (2) close follow-up of such patients by clinical and two-dimensional echocardiographic evaluation is advisable after the third postoperative year, and (3) failure of the mitral Ionescu-Shiley pericardial xenograft may occur suddenly, and awareness of this complication is the clue to prompt recognition and treatment of such patients.

摘要

在1977年1月至1980年12月期间,40例接受Ionescu-Shiley心包异种移植物二尖瓣置换术出院的患者中,发生了7例瓣膜功能衰竭。Ionescu-Shiley心包异种移植物二尖瓣置换术后生物假体功能障碍的这种前所未有的发生率促使我们更新对二尖瓣位植入该瓣膜患者的随访。我们的调查结果显示,术后6年,精算生存率为72%,无栓塞的精算概率为62%,无假体功能衰竭的精算概率为60%。取出的瓣膜显示,涤纶覆盖框架完全没有新内膜衬里,且存在与我们疲劳测试系统中平均经过29±17×10⁶次循环后取出的Ionescu-Shiley心包异种移植物中观察到的病变相似的病变。这种相似性促使对二尖瓣Ionescu-Shiley心包异种移植物中出现的撕裂进行分类,这极为重要,因为瓣膜功能衰竭患者的临床表现和结局因破裂类型而异。临床标本中观察到的瓣叶病变可能与瓣膜关闭时组织与裸露的涤纶布持续摩擦有关。结论是:(1)Ionescu-Shiley心包异种移植物在二尖瓣位的实际耐久性和血栓形成性需要仔细重新评估;(2)术后第三年之后,建议通过临床和二维超声心动图评估对这类患者进行密切随访;(3)二尖瓣Ionescu-Shiley心包异种移植物可能突然发生功能衰竭,认识到这种并发症是及时识别和治疗这类患者的关键。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验