Owen-Falkenberg A P, Møller J, Owen-Falkenberg T
Acta Anaesthesiol Scand. 1984 Oct;28(5):576-7. doi: 10.1111/j.1399-6576.1984.tb02122.x.
The variation in micro-drop size was studied for a series of sedative-anaesthetic solutions administered via a standard paediatric infusion set in combination with an infusion pump and drop counter. Of the solutions tested, only Diazemuls 20 mg% failed to alter micro-drop size. Solutions containing Ketalar, Apozepam, Hypnomidate and Althesin all caused a decrease in drop size and therefore in administered volume. The largest error (37%) occurred with a solution containing althesin 20% (v/v).