Suppr超能文献

透视控制下经皮肾镜碎石术中的辐射暴露。

Radiation exposure during fluoroscopically controlled percutaneous lithotripsy.

作者信息

Kelsey C A, Lane R G, Somers J W

出版信息

J Urol. 1984 Dec;132(6):1254-5. doi: 10.1016/s0022-5347(17)50117-2.

Abstract

Radiation measurements in the vicinity of a standard urologic table modified for fluoroscopic-guided lithotripsy demonstrate radiation levels as high as 3000 mrem per hour at the edge of the table. Excessive radiation exposure to the physician, exceeding the annual maximum permissible dose to the eyes and thyroid, can be expected after only 10 cases. Suggested radiation protection measures include the use of finger, wrist and neck radiation monitors, eye and thyroid shields and improved shielding of both collimator and patient scatter radiation.

摘要

在为荧光透视引导碎石术改装的标准泌尿科检查台附近进行的辐射测量显示,检查台边缘的辐射水平高达每小时3000毫雷姆。仅10例手术后,医生就可能受到超过眼睛和甲状腺年度最大允许剂量的过量辐射照射。建议的辐射防护措施包括使用手指、手腕和颈部辐射监测器、眼睛和甲状腺防护屏,以及改进准直器和患者散射辐射的屏蔽。

相似文献

2
Radiation exposure to patient and urologist during percutaneous nephrostolithotomy.
J Urol. 1984 Dec;132(6):1148-52. doi: 10.1016/s0022-5347(17)50071-3.
6
Occupational radiation exposure to the surgeon.外科医生的职业性辐射暴露。
J Am Acad Orthop Surg. 2005 Jan-Feb;13(1):69-76. doi: 10.5435/00124635-200501000-00009.