一种用于血清中茶碱酶免疫测定的全自动化系统。

A totally automated system for enzyme immunoassay of theophylline in serum.

作者信息

Castro A, Ibanez J, Voight W, Noto T, Malkus H

出版信息

Clin Chem. 1978 Jun;24(6):944-6.

PMID:657486
Abstract

We describe a procedure for enzyme immunoassay of theophylline (1,3-dimethylxanthine) in which all phases of the assay are totally automated in a Kinetic Analyzer (KA-150). This system permits assay of 75 10-mul samples per hour, with results available at 30-s intervals after initial sample preparation and preincubation. We compared results for 138 clinical samples by an ultraviolet method (x) and the present method (y). The slope of the comparison curve was 0.902, the y-intercept 0.402, and the correlation coefficient 0.984. The coefficient of variation for samples run in duplicate on the same day was 4.9%; it was 8.1% for samples run on different days. Specificity, sensitivity, simplicity, speed, and small reagent requirement all make this an attractive alternative to chromatographic procedures.

摘要

我们描述了一种在动力学分析仪(KA - 150)中对茶碱(1,3 - 二甲基黄嘌呤)进行酶免疫测定的方法,该方法的所有测定阶段均可在其中完全自动化。该系统每小时可测定75个10微升的样本,在初始样本制备和预孵育后,每隔30秒即可获得结果。我们通过紫外法(x)和本方法(y)对138份临床样本的结果进行了比较。比较曲线的斜率为0.902,y轴截距为0.402,相关系数为0.984。同一天重复检测样本的变异系数为4.9%;不同天检测样本的变异系数为8.1%。特异性、灵敏度、简便性、速度以及试剂需求量少等特点,使得该方法成为色谱法颇具吸引力的替代方法。