Leake-Jones P A, Rebscher S J
Ann N Y Acad Sci. 1983;405:203-23. doi: 10.1111/j.1749-6632.1983.tb31634.x.
In summary, these results with chronic implantation of three types of scala tympani electrode indicate the critical importance of two factors in reducing the risk of additional damage to cochlear structure by implantation surgery: (1) The shape and mechanical characteristics of the array must be precisely controlled such that insertion can be performed with an acceptably low incidence of trauma; and (2) the specific electrode materials and fabrication procedures must be demonstrated to be highly biocompatible in in vivo animal control studies. It appears that the neuronal elements of a prior normal cochlea (in the cat) can withstand chronic implantation of scala tympani electrodes (for at least 1 year) if these two prerequisites are met. The effects of chronic electrical stimulation with these arrays at current levels appropriate for the operation of such devices in patients is not known at present.
总之,三种类型的鼓阶电极长期植入的这些结果表明,在降低植入手术对耳蜗结构造成额外损伤风险方面,有两个因素至关重要:(1)阵列的形状和机械特性必须精确控制,以便在创伤发生率可接受的情况下进行插入;(2)在体内动物对照研究中,必须证明特定的电极材料和制造工艺具有高度生物相容性。如果满足这两个先决条件,先前正常耳蜗(猫的)神经元成分似乎能够承受鼓阶电极的长期植入(至少1年)。目前尚不清楚以适合此类装置在患者中运行的电流水平,用这些阵列进行长期电刺激的效果如何。