Sabin G, Schnieder B, Neuhausen P, Berkmen R
Dtsch Med Wochenschr. 1978 Jun 16;103(24):1012-5. doi: 10.1055/s-0028-1104823.
For the assessment of stable positions of the electrodes radiographic control, threshold measurements and methods provoking dislocation are used. There is an average dislocation rate of endocardial electrodes of 7-11%. As a further parameter for intra-operative assessment of a stable electrode position intracardiac R-potentials were measured with the stimulation electrode. While QRS complexes in the surface ECG remain unchanged, intracardiac electrocardiograms show an alternating height of the R-amplitude from one ventricular action to the next. From these differences (deltaR) a stable electrode position can be assumed as long as deltaR is small (our patient group 2.12mV+/-1.05MV). Large differences (6.9mV+/-1.7MV) indicate a non-optimal fixation of the electrode tip. The dislocation rate including the post-operative exit block could be reduced below 2% by consideration of the R-amplitude alterations.
为评估电极的稳定位置,采用了X线控制、阈值测量及诱发脱位的方法。心内膜电极的平均脱位率为7%-11%。作为术中评估电极稳定位置的另一参数,用刺激电极测量心腔内R波电位。体表心电图的QRS波群保持不变,而心腔内心电图显示,从一个心室动作到下一个心室动作,R波振幅高度交替变化。只要差值R较小(我们的患者组为2.12mV±1.05MV),就可认为电极位置稳定。差值R较大(6.9mV±1.7MV)表明电极尖端固定不佳。通过考虑R波振幅变化,包括术后出口阻滞在内的脱位率可降至2%以下。