Sabin G, Schnieder B
Herz. 1980 Aug;5(4):257-61.
The dislodgement rate of transvenous, endocardial pacemaker leads ranges between 7 and 11%. To evaluate the stability of the lead position, intraoperative measurements of the intracardial R-potential were performed via the positioned stimulating electrode. The intracardial R-potential demonstrated fluctuating amplitudes while those of the simultaneously recorded surface ECG remained constant. The fluctuations in amplitude were measured as delta-R. In the presence of only slight fluctuations in R-amplitude (delta-R 2.10 mV, SD = +/- 1.08 mV) the position of 250 electrodes was documented to be securely fixed. In the case of 23 electrodes with marked fluctuations in R-amplitude (delta R 6.68 +/- 1.7 mV) intraoperative dislocation occurred in spite of adequate positioning on fluoroscopy and low thresholds. In five patients failure of the pacing function occurred postoperatively; in this group the mean measured value of delta-R was 5.66 mV. Particularly small fluctuations in delta-R were observed in association with cork-screw electrodes and those which had been stabily positioned for a number of years; the mean value for the latter groups was 0.7 +/- 0.3 mV. The uniformity of the intracardial R-amplitude is primarily dependent on the mechanical stability of the electrode tip. Even slight geometrical changes may lead to notable amplitude fluctuations. In this study, measurement of the R-amplitude fluctuations to guide positioning of the electrode permitted a reduction of the dislodgement rate to 2%. Technically, the method does not require complicated instrumentation and is not time-consuming.
经静脉心内膜起搏器导线的脱位率在7%至11%之间。为评估导线位置的稳定性,通过已定位的刺激电极在术中测量心内R电位。心内R电位的幅度波动不定,而同时记录的体表心电图的幅度保持恒定。幅度波动以ΔR来衡量。当R幅度仅有轻微波动(ΔR 2.10 mV,标准差=±1.08 mV)时,记录显示250根电极的位置固定牢固。在23根电极R幅度有明显波动(ΔR 6.68±1.7 mV)的情况下,尽管在荧光透视下定位良好且阈值较低,但术中仍发生了脱位。5例患者术后起搏功能失败;该组中ΔR的平均测量值为5.66 mV。在与螺旋电极以及已稳定放置数年的电极相关的情况下,观察到ΔR的波动特别小;后一组的平均值为0.7±0.3 mV。心内R幅度的均匀性主要取决于电极尖端的机械稳定性。即使是轻微的几何形状变化也可能导致明显的幅度波动。在本研究中,通过测量R幅度波动来指导电极定位,可将脱位率降低至2%。从技术上讲,该方法不需要复杂的仪器设备,也不耗时。