Bain B, Dean A, Broom G
Clin Lab Haematol. 1984;6(3):273-85. doi: 10.1111/j.1365-2257.1984.tb00553.x.
The accuracy of the Coulter Counter Model S Plus III in determining lymphocyte percentage was assessed. In 'routine' adult blood counts accuracy was satisfactory but in infectious mononucleosis, and in chronic lymphocytic leukaemia and other lymphoproliferative disorders, there was a high rejection rate and an often considerable underestimation of the lymphocyte percentage. The S Plus III lymphocyte percentage was of no use in neonates because of inaccuracy and a high rejection rate. In infants the relationship with lymphocyte percentages based on manual differential counts was non-linear, and there was underestimation of some high percentages. The white cell size histogram was useful in altering staff to abnormalities, and also suggested that lymphocytes of neonates and infants were larger than those of adults, with plots being very similar to those of some adults with lymphoproliferative disorders. The S Plus III lymphocyte percentage was useful in reducing the need for differential counts in adult patients but did not replace examination of a blood film or a manual lymphocyte percentage in patients with chronic lymphocytic leukaemia or a lymphoproliferative disorder.
对库尔特S Plus III型血细胞计数器测定淋巴细胞百分比的准确性进行了评估。在“常规”成人血细胞计数中,准确性令人满意,但在传染性单核细胞增多症、慢性淋巴细胞白血病和其他淋巴细胞增生性疾病中,拒收率很高,且淋巴细胞百分比常常被严重低估。由于不准确和高拒收率,S Plus III型血细胞计数器测定的淋巴细胞百分比在新生儿中毫无用处。在婴儿中,其与基于手工分类计数得出的淋巴细胞百分比之间的关系是非线性的,并且一些高百分比被低估。白细胞大小直方图有助于工作人员发现异常情况,还表明新生儿和婴儿的淋巴细胞比成人的大,其图谱与一些患有淋巴细胞增生性疾病的成人非常相似。S Plus III型血细胞计数器测定的淋巴细胞百分比有助于减少成人患者分类计数的需求,但并不能取代对慢性淋巴细胞白血病或淋巴细胞增生性疾病患者的血涂片检查或手工淋巴细胞百分比测定。