对库尔特S Plus IV型血细胞分析仪上的三分类血细胞计数进行评估。

An assessment of the three-population differential count on the Coulter Counter Model S Plus IV.

作者信息

Bain B J

出版信息

Clin Lab Haematol. 1986;8(4):347-59. doi: 10.1111/j.1365-2257.1986.tb00123.x.

Abstract

The accuracy and precision of a three-population differential count on a Coulter Counter S Plus IV were assessed. In general a good correlation was found between manual and automated lymphocyte and granulocyte percentages although various abnormal samples caused discrepancies, and it was found that some eosinophils were categorized as mononuclear cells rather than granulocytes. The automated mononuclear cell percentage was found to be imprecise and often inaccurate as a reflection of monocyte numbers. The instrument sometimes failed to either reject or 'flag' samples containing an appreciable percentage of lymphoma cells, nucleated red blood cells, blast cells, or eosinophils but was found useful in reducing the number of differential counts required.

摘要

对库尔特计数器S Plus IV上三群细胞分类计数的准确性和精密度进行了评估。总体而言,尽管各种异常样本导致了差异,但手动和自动检测的淋巴细胞及粒细胞百分比之间存在良好的相关性,并且发现一些嗜酸性粒细胞被归类为单核细胞而非粒细胞。发现自动检测的单核细胞百分比不准确,不能准确反映单核细胞数量。该仪器有时无法拒收或“标记”含有相当比例淋巴瘤细胞、有核红细胞、原始细胞或嗜酸性粒细胞的样本,但发现其有助于减少所需的分类计数数量。