Rosenbaum T J, Sundt T M
J Neurosurg. 1978 Jun;48(6):929-34. doi: 10.3171/jns.1978.48.6.0929.
Various straight-jawed aneurysm clips were tested for occluding capabilities on a vascular tissue model. Occluding pressures varied markedly among the clip styles and were altered by changes in the lumen and tissue composition of the model. Mechanical characteristics are highly variable between clip styles, but fall within a narrow range for clips of a similar style. The complex interplay of the unique aspects of clip design and force generated by the spring in conjunction with tissue characteristics and precise clip placement upon the tissue are major variables in the apparent ability of a particular clip to occlude the neck of an aneurysm.
在血管组织模型上对各种直钳口动脉瘤夹的闭塞能力进行了测试。不同类型的夹子之间,闭塞压力差异显著,且会因模型管腔和组织成分的变化而改变。不同类型夹子的机械特性差异很大,但同一类型的夹子其特性范围较窄。夹子设计的独特方面、弹簧产生的力、组织特性以及夹子在组织上的精确放置之间复杂的相互作用,是特定夹子闭塞动脉瘤颈部明显能力的主要变量。