Vogel R A, Kirch D, LeFree M, Steele P
J Nucl Med. 1978 Jun;19(6):648-54.
A new method of multiplanar emission tomography is described; it uses a wide-field Anger scintillation camera (37.5 cm crystal diameter) and a seven-pinhole collimator. The pinholes (5.5 mm) acquire data simultaneously from the emissing source and project the data onto seven independent regions of the camera crystal. Multiple planes are reconstructed from the initial seven-view data acquisition through the use of a computerized addition-multiplication algorithm and variation of the superposition relationships among the projected views. These planes are then altered iteratively by a least-error criterion following ray-sum comparison with the original views. Planar resolution (full-width-half-maximum) is 1.0 cm and depth resolution is 1.5 cm. In 42 patients with angiographically demonstrated coronary-artery disease, studies of myocardial Tl-201 perfusion, under exercise, have shown improved detection sensitivity in comparison with scintigraphy using parallel-hole collimation.
本文描述了一种新的多平面发射断层扫描方法;它使用宽视野安格尔闪烁相机(晶体直径37.5厘米)和七针孔准直器。针孔(5.5毫米)同时从发射源获取数据,并将数据投射到相机晶体的七个独立区域。通过使用计算机加法-乘法算法以及改变投影视图之间的叠加关系,从最初的七视图数据采集中重建多个平面。然后,通过与原始视图进行射线和比较,按照最小误差标准对这些平面进行迭代更改。平面分辨率(半高宽)为1.0厘米,深度分辨率为1.5厘米。在42例经血管造影证实患有冠状动脉疾病的患者中,对运动状态下心肌Tl-201灌注的研究表明,与使用平行孔准直的闪烁扫描相比,其检测灵敏度有所提高。