Steinfeld L, Dimich I, Reder R, Cohen M, Alexander H
J Pediatr. 1978 Jun;92(6):934-8. doi: 10.1016/s0022-3476(78)80364-3.
The indirect measurement of blood pressure in the pediatric population is often difficult or impossible to perform and when values are obtained, they are often unreliable or inaccurate. An accurate measurement of arterial systolic pressure in the upper extremity is obtainable with a properly designed and sized occluding cuff and an ultrasonic device to detect the first Korotoff sound. For the detection of the first Korotkoff sound ultrasonic Doppler devices are superior to the stethoscope. When wider occluding cuffs than usual are applied to the upper extremity, the error of measurement is minimized. A method for accurately measuring end diastolic pressure has yet to be described despite assertions to the contrary.
在儿科人群中,间接测量血压往往很难进行或根本无法进行,即便获得了血压值,这些值通常也不可靠或不准确。使用设计合理且尺寸合适的袖带和超声设备来检测第一声柯氏音,能够准确测量上肢动脉收缩压。在检测第一声柯氏音方面,超声多普勒设备优于听诊器。将比平常更宽的袖带用于上肢时,测量误差可降至最低。尽管有相反的说法,但准确测量舒张末期压力的方法尚未见报道。