Pridgar E M, Leung F Y, Henderson A R
Clin Chem. 1983 Dec;29(12):2096-9.
We have further assessed the accuracy of the thin-layer agarose fluorescent technique of Elevitch et al. [Am J Clin Pathol 46, 692 (1966)]. Previously, we used semi-purified human lactate dehydrogenase isoenzyme-1 and -5 [Clin Chem 22, 1995 (1976)] and isoenzyme-1 and -2 [Clin Chem 27, 1708 (1981)] to show that this assay accurately measures the proportions of these binary mixtures. In the present study, using ternary and quaternary mixtures of isoenzyme- 1, -2, -3, and -5, we show that the assay gives accurate estimations of all of these isoenzymes, within the errors of the techniques used. We also show that peak area (integration) is more nearly accurate, but less precise, than peak height (amplitude) measurements.
我们进一步评估了Elevitch等人[《美国临床病理学杂志》46, 692 (1966)]的薄层琼脂糖荧光技术的准确性。此前,我们使用半纯化的人乳酸脱氢酶同工酶-1和-5[《临床化学》22, 1995 (1976)]以及同工酶-1和-2[《临床化学》27, 1708 (1981)]来表明该测定法能准确测量这些二元混合物的比例。在本研究中,我们使用同工酶-1、-2、-3和-5的三元和四元混合物,表明在所用技术的误差范围内,该测定法能对所有这些同工酶进行准确估计。我们还表明,峰面积(积分)比峰高(幅度)测量更接近准确,但精度较低。