Klein D J, Chan H P, Muntz E P, Doi K, Lee K, Chopelas P, Bernstein H, Lee J
Med Phys. 1983 Sep-Oct;10(5):664-8. doi: 10.1118/1.595369.
Predicted scatter characteristics from a Monte Carlo code have been compared to an experiment for a 4.0-cm-thick acrylic resin (Plexiglas) phantom and 27.3-keV radiation. The results show generally good agreement, except that in the scattered spectra the predictions appear to underestimate the proportion of photons in the low-energy tail of the scattered radiation. The angular dependency of the energy-integrated scattered radiation fluence is in excellent agreement with the predictions.
已将蒙特卡罗代码预测的散射特性与针对4.0厘米厚的丙烯酸树脂(有机玻璃)体模和27.3千电子伏辐射进行的实验进行了比较。结果总体显示出良好的一致性,只是在散射光谱中,预测结果似乎低估了散射辐射低能尾部光子的比例。能量积分散射辐射注量的角度依赖性与预测结果高度吻合。