Frantzen Handeland G, Abildgaard U, Aasen A O
Scand J Haematol. 1983 Nov;31(5):427-36. doi: 10.1111/j.1600-0609.1983.tb01538.x.
A simplified method for the assay of antithrombin III (AT) with the highly reactive thrombin substrate 2AcOH X H-D-CHG-Ala-Arg-pNA (substrate Th-1) is described. The assay may be performed at either 30 degrees C or 37 degrees C, and alternatively with the substrate H-D-Phe-Pip-Arg-pNA (S-2238). The standard curve is linear in the 12.5-150% range. For routine assays, 3 standard dilutions of plasma are sufficient, and these may be stored at -20 degrees C for 3 weeks. As only the test plasma must be diluted prior to the assay procedure, the test is more rapidly performed than previous manual assays. In 80 patients plasma samples, with AT in the 19-108% range, there was a high correlation with the results of immunoquantification (r = 0.96). There was also a high correlation between the results obtained with the manual method and the automated version described using the Cobas-Bio Centrifugal Analyser and substrate Th-1 (r = 0.96). Low AT levels in hereditary deficiency (particularly during heparin treatment), in liver cirrhosis, in disseminated intravascular coagulation (DIC), and heparin-treated thrombosis were confirmed.
描述了一种使用高反应性凝血酶底物2AcOH X H-D-CHG-Ala-Arg-pNA(底物Th-1)测定抗凝血酶III(AT)的简化方法。该测定可在30℃或37℃下进行,也可使用底物H-D-Phe-Pip-Arg-pNA(S-2238)。标准曲线在12.5 - 150%范围内呈线性。对于常规测定,血浆的3种标准稀释液就足够了,并且这些稀释液可在-20℃下保存3周。由于仅测试血浆在测定程序前必须稀释,所以该测试比以前的手工测定执行得更快。在80例血浆样本中,AT水平在19 - 108%范围内,与免疫定量结果高度相关(r = 0.96)。使用手工方法获得的结果与使用Cobas-Bio离心分析仪和底物Th-1描述的自动化版本获得的结果之间也高度相关(r = 0.96)。遗传性缺乏(特别是在肝素治疗期间)、肝硬化、弥散性血管内凝血(DIC)和肝素治疗的血栓形成中的低AT水平得到了证实。