Abstract

The accuracy of the spectral difference method for measuring ultrasonic attenuation has been investigated using tissue-mimicking phantoms. Attenuation coefficients of the phantom materials were measured using a narrow-band substitution technique and compared with the results of the spectral difference method. Agreement within +/-10 percent was typical for measurements in homogeneous materials. The best agreement between the spectral difference and substitution techniques was obtained when effects due to transducer beam diffraction were taken into account in the analysis. This was found for two types of homogeneous tissue-mimicking materials, both having speed of sound and attenuation properties similar to human liver but each with different backscatter properties. The effects of inhomogeneous tissues interposed between the transducer and the interrogated volume were also studied by simulating these conditions in phantoms. Experimental techniques which minimize the effects of perturbations introduced by these inhomogeneities are suggested.

摘要

已使用仿组织体模研究了用于测量超声衰减的光谱差分法的准确性。使用窄带替代技术测量了体模材料的衰减系数,并与光谱差分法的结果进行了比较。在均匀材料中的测量通常在±10%的范围内一致。当在分析中考虑换能器波束衍射的影响时,光谱差分法和替代技术之间获得了最佳一致性。在两种类型的均匀仿组织材料中发现了这一点,这两种材料的声速和衰减特性均与人类肝脏相似,但每种材料具有不同的背散射特性。还通过在体模中模拟这些条件研究了置于换能器和被询问体积之间的不均匀组织的影响。提出了将这些不均匀性引入的扰动影响降至最低的实验技术。