Lam K S, Taborsky C, Lam W C
Comput Biomed Res. 1983 Apr;16(2):169-75.
An on-line data acquisition system for external beam treatment planning has been developed. The hardware consists of a Tektronix 4054 Computer, an SHM Nuclear 3-D scanning water phantom, interfaces for 16 channels of ADC and 32 channels of digital I/O, and two detector probes for measuring dose distribution and monitoring fluctuations in machine output. The program to control the movement of the scanning probe and to collect data is written in BASIC. Results of the measurements are plotted in real time on the Tektronix screen. The beam data format consists of a set of percentage depth doses along the central axis and five sets of off-axis ratios at five different depths. In the electron mode, when ionization chambers are used, data points are averages of two measurements, one with positive bias and the other with negative bias. The beam data are stored on disk which can be assessed by the Capintec treatment planning system.
已开发出一种用于外照射治疗计划的在线数据采集系统。硬件包括一台泰克4054计算机、一个SHM Nuclear三维扫描水模体、用于16通道模数转换器和32通道数字输入/输出的接口,以及两个用于测量剂量分布和监测机器输出波动的探测器探头。控制扫描探头移动和收集数据的程序用BASIC语言编写。测量结果实时绘制在泰克屏幕上。射束数据格式由一组沿中心轴的百分深度剂量和五组在五个不同深度的离轴比组成。在电子模式下,当使用电离室时,数据点是两次测量的平均值,一次是正偏置,另一次是负偏置。射束数据存储在磁盘上,可由Capintec治疗计划系统进行评估。