Stepanik J
Klin Monbl Augenheilkd. 1984 Jan;184(1):40-1. doi: 10.1055/s-2008-1054406.
The scale-reading of the Goldmann tonometer (1 g mass = 9.81 mN force) corresponds to an intraocular pressure of 10 mm Hg at a corneal applanation diameter of d = 3.06 mm (applanating body 1, x values) and to an intraocular pressure of 1 kPa at a corneal applanation diameter of d = 3.53 mm (applanating body 2, y values). A regression line y, calculated from 100 pairs of measurements with these two applanating bodies, with y = 0.097 + 0.1274x and a standard deviation of the y values of Syx = +/- 0.24, was within the 95% limits of tolerance at all points. Hence, the two applanating bodies are equally acceptable.
戈德曼眼压计的刻度读数(1克质量 = 9.81毫牛顿力)在角膜压平直径d = 3.06毫米时(压平体1,x值)对应于10毫米汞柱的眼压,在角膜压平直径d = 3.53毫米时(压平体2,y值)对应于1千帕的眼压。由使用这两个压平体的100对测量值计算出的回归线y,其方程为y = 0.097 + 0.1274x,y值的标准差Syx = +/- 0.24,在所有点上均处于95%的公差范围内。因此,这两个压平体同样适用。