0.5特斯拉磁共振成像中的呼吸门控

Respiratory gating in magnetic resonance imaging at 0.5 Tesla.

作者信息

Runge V M, Clanton J A, Partain C L, James A E

出版信息

Radiology. 1984 May;151(2):521-3. doi: 10.1148/radiology.151.2.6709928.

Abstract

A device for gating the acquisition of magnetic resonance images with chest wall motion was developed, and the effects of respiratory gating upon image quality were studied. Images of respiratory gated examinations were compared with those of ungated examinations in 16 subjects. In a subset of four of those subjects, combined respiratory and cardiac gated images were obtained. Respiratory gating removed gross motion artifacts from magnetic resonance images of the chest and abdomen. Resolution of small normal tissue structures, such as the portal and hepatic veins, is improved. In cardiac studies, respiratory gating improved the visualization and definition of the atria and ventricles.

摘要

研发了一种用于根据胸壁运动对磁共振图像采集进行门控的设备,并研究了呼吸门控对图像质量的影响。对16名受试者的呼吸门控检查图像与非门控检查图像进行了比较。在其中4名受试者的子集中,获得了呼吸和心脏联合门控图像。呼吸门控消除了胸部和腹部磁共振图像中的明显运动伪影。门静脉和肝静脉等小的正常组织结构的分辨率得到了提高。在心脏研究中,呼吸门控改善了心房和心室的可视化及清晰度。