Björk-Shiley瓣膜假体的晚期支柱骨折(当前系列)

Late strut fracture in a Björk-Shiley valve prosthesis (current series).

作者信息

Sacks S H, Northeast A D, Watkins J

出版信息

Br Heart J. 1984 May;51(5):578-80. doi: 10.1136/hrt.51.5.578.

Abstract

Fracture of the short strut of the latest series of Björk-Shiley mitral valve prostheses occurred 12 months after implantation. The patient died despite emergency surgery to excise and replace the prosthesis. Björk-Shiley 29-33 mm valves with a 60 degree opening angle manufactured between February 1981 and March 1982 have a predicted strut fracture rate of 0.3% (three times greater than that of valves produced outside these dates). Unimplanted valves from this batch have now been recalled by the manufacturers.