Suppr超能文献

超声组织纹理:换能器聚焦模式的影响

Sonographic tissue texture: influence of transducer focusing pattern.

作者信息

Jaffe C C, Harris D J

出版信息

AJR Am J Roentgenol. 1980 Aug;135(2):343-7. doi: 10.2214/ajr.135.2.343.

Abstract

The ultrasonic beam profile pattern is not accurately known for most transducers in clinical use. A simple scan of an easily available phantom permits the clinician to determine the focusing pattern of his own transducer and equipment. Disregarding other operator-controllable parameters such as the time-gain compensation, output, and system sensitivity, the true echo pattern of a tissue is not accurately represented until the distance of the region of narrowest beam focus is reached. The fine speckled echo pattern commonly seen in uniform organs near the transducer is largely artifactual due to interference phenomena in the near field of the acoustic beam. Closest approximation of true texture is not displayed until near the depth at which the narrowest focus occurs. Beyond that depth, broad, coarser patterns can be expected.

摘要

对于临床使用的大多数换能器而言,超声束轮廓模式并不确切知晓。对一个容易获得的体模进行简单扫描,可使临床医生确定其自身换能器和设备的聚焦模式。忽略其他可由操作员控制的参数,如时间增益补偿、输出和系统灵敏度,直到达到波束最窄聚焦区域的距离时,组织的真实回声模式才会被准确呈现。由于声束近场中的干涉现象,在换能器附近均匀器官中常见的精细斑点状回声模式在很大程度上是人为的。直到接近出现最窄聚焦的深度时,才会显示出最接近真实纹理的图像。在该深度之后,可以预期会出现更宽、更粗糙的模式。