The concentration of IgA, IgG, IgM, a1-antitrypsin, haptoglobin and transferrin was determined by the kinetic nephelometric method (Beckman immunochemistry analyzer) and by the radial immunodiffusion method (Behring partigen immunodiffusion plates). The investigation was carried out in 151 normal and pathological sera. The results obtained with the radial immunodiffusion method were for IgA, IgG and IgM in middle about 15%, for a1-antitrypsin about 50% and for haptoglobin about 80% higher as compared to the results determined with the kinetic nephelometric method. An acceptable correspondence between the two methods was found for the results of transferrin.
采用动力学散射比浊法(贝克曼免疫化学分析仪)和放射免疫扩散法(贝林免疫扩散板)测定了IgA、IgG、IgM、α1-抗胰蛋白酶、触珠蛋白和转铁蛋白的浓度。对151份正常和病理血清进行了检测。与动力学散射比浊法测定的结果相比,放射免疫扩散法测定的IgA、IgG和IgM结果在中等水平约高15%,α1-抗胰蛋白酶约高50%,触珠蛋白约高80%。两种方法测定的转铁蛋白结果具有可接受的一致性。