Suppr超能文献

激光散射比浊法测定七种血清蛋白并与放射免疫扩散法比较。

Laser nephelometric measurement of seven serum proteins compared with radial immunodiffusion.

作者信息

Guiguet M, Padieu P, Mack G, Dalebroux M

出版信息

J Clin Chem Clin Biochem. 1983 Apr;21(4):217-21. doi: 10.1515/cclm.1983.21.4.217.

Abstract

Quantification of immunoglobulins IgG, IgA, IgM, C3 complement, siderophilin, alpha 2-macroglobulin and haptoglobin in serum by a Behring laser nephelometer coupled with a date processing apparatus was compared with values obtained by the radial immunodiffusion method of Mancini. The precision was not improved by the use of the laser nephelometer as compared to Mancini radial immunodiffusion, but within run precision was better than that among runs for both methods. The analytical recovery studies for laser nephelometry showed a close concordance between the theoretical and measured values for all proteins. The two techniques were also compared by means of correlation studies.

摘要

采用贝林激光散射比浊仪结合数据处理装置对血清中的免疫球蛋白IgG、IgA、IgM、补体C3、转铁蛋白、α2-巨球蛋白和触珠蛋白进行定量分析,并与通过曼奇尼径向免疫扩散法获得的值进行比较。与曼奇尼径向免疫扩散法相比,使用激光散射比浊仪并未提高精密度,但两种方法的批内精密度均优于批间精密度。激光散射比浊法的分析回收率研究表明,所有蛋白质的理论值和测量值之间具有密切的一致性。还通过相关性研究对这两种技术进行了比较。