Seven-electrode EER stimulator-monitor for ambulatory monkeys.

作者信息

Deutsch S

出版信息

Biotelem Patient Monit. 1980;7(3-4):231-41.

PMID:6783145
Abstract

An electrical evoked response (EER) stimulator-monitor for ambulatory monkeys is described. Seven time-multiplexed electrodes are used. The unit switches between stimulus and monitor modes at a 100-Hz rate. In the stimulus mode, 3.5 V peak, 0.45-ms pulses are applied at a 100-Hz rate to any electrode selected by means of toggle switches on a remote control panel. In the monitor mode, low-frequency -3 dB cutoff occurs at 0.2 Hz. A 20-MHz carrier is used as a local superheterodyne oscillator during the stimulus period, and as an FM transmitter during the monitor period. Three mercury cells supply 4 V at 10 mA. The unit weighs 80 g. It is 5.7 x 5 x 2.5 cm in size.

摘要

描述了一种用于动态监测猴子的电诱发反应(EER)刺激监测器。使用了七个时分复用电极。该装置以100赫兹的速率在刺激模式和监测模式之间切换。在刺激模式下,通过远程控制面板上的拨动开关选择任何一个电极,以100赫兹的速率施加3.5伏峰值、0.45毫秒的脉冲。在监测模式下,低频-3分贝截止频率为0.2赫兹。在刺激期,使用20兆赫兹载波作为本地超外差振荡器,在监测期作为调频发射机。三个汞电池以10毫安的电流提供4伏电压。该装置重80克,尺寸为5.7×5×2.5厘米。