Meyers A
Ann Otol Rhinol Laryngol. 1981 Mar-Apr;90(2 Pt 1):132-4. doi: 10.1177/000348948109000207.
Since the introduction of CO2 laser surgery, surgeons have stressed the importance of protecting nontarget structures from errant irradiation. Although striking areas outside the intended field may inflict little or no damage, irradiating an exposed endotracheal tube may result in an endotracheal fire. Recent experience with a case of endotracheal tube ignition led to reevaluation of those factors associated with this serious complication. In an attempt to determine the best endotracheal tube to use during laser surgery of the larynx, the flammability of polyvinyl chloride and red rubber endotracheal tubes was compared. Several factors were noted to contribute to the potential hazard of endotracheal tube ignition. Since red rubber tubes are less flammable, they should be used during laryngeal CO2 laser surgery.
自二氧化碳激光手术引入以来,外科医生一直强调保护非目标结构免受误照射的重要性。尽管照射预期区域之外的部位可能造成很小的损害或没有损害,但照射暴露的气管内导管可能会引发气管内火灾。最近一例气管内导管着火的病例促使人们重新评估与这种严重并发症相关的因素。为了确定在喉部激光手术中使用的最佳气管内导管,对聚氯乙烯和红色橡胶气管内导管的可燃性进行了比较。人们注意到有几个因素会导致气管内导管着火的潜在危险。由于红色橡胶导管不易燃,在喉部二氧化碳激光手术中应使用此类导管。