Treyve E, Yarington C T, Thompson G E
Ann Otol Rhinol Laryngol. 1981 Jul-Aug;90(4 Pt 1):328-30. doi: 10.1177/000348948109000407.
The ignition characteristics of the American Hospital, Portex, polyvinyl chloride, and red rubber tracheal tubes commonly available have been evaluated using a standard carbon dioxide laser. The added hazard of the increased combustibility of the impregnated line on endotracheal tubes and the hazard of commonly utilized anesthetic or lubricating materials are discussed. The safety limits which experimentally, indicate levels of energy and laser burst time which seem to preclude the hazard of ignition are outlined. The potential danger of postburn bronchiolitis is noted, along with appropriate treatment.
已使用标准二氧化碳激光评估了市面上常见的美国医院用、波特克斯(Portex)、聚氯乙烯和红色橡胶气管导管的着火特性。讨论了气管导管上浸渍线可燃性增加带来的额外危害以及常用麻醉或润滑材料的危害。概述了通过实验得出的安全限值,这些限值表明了似乎可避免着火危险的能量水平和激光脉冲时间。还指出了烧伤后细支气管炎的潜在危险以及相应的治疗方法。