新的气液平衡方法:注射器压力计。

New gas-liquid equilibration method: syringe tonometer.

作者信息

Wallace W D, Cutler C A, Clark J S

出版信息

Clin Chem. 1981 May;27(5):681-6.

PMID:6784964
Abstract

This new apparatus for gas-liquid equilibration (tonometry) in a transportable vessel is designed for tonometry of blood or buffer solution in a specially designed syringe. Gas enters the syringe chamber through small holes in the tip of the syringe plunger and bubbles upward through the sample. The syringe plunger is a second chamber, which is used for warming and humidifying the gas before it enters the tonometer chamber. The entire syringe is housed in a transparent, temperature-controlled environment during equilibration. After equilibration, the sample is easily entered into a blood-gas analyzer. At most, gas-liquid O2/CO2 equilibration for 2.5 mL of buffer or blood requires less than 13 min. Comparisons with a standard thin-film tonometer show good agreement for pO2 and pCO2 over the range 0-93 kPa (0-700 mmHg) and 2-20 kPa (14-150 mmHg), respectively. The syringe tonometer eliminates contamination of the sample during transfer and transport to the analyzer, thus making tonometry more technique-independent. The apparatus is simple and easy to use, with definite advantages over existing methods of tonometry.

摘要

这种用于在可移动容器中进行气液平衡(测压法)的新装置,是为在特制注射器中对血液或缓冲溶液进行测压而设计的。气体通过注射器柱塞尖端的小孔进入注射器腔室,并向上穿过样品形成气泡。注射器柱塞是第二个腔室,用于在气体进入测压腔室之前对其进行加热和加湿。在平衡过程中,整个注射器置于透明的、温度可控的环境中。平衡后,样品可轻松进入血气分析仪。对于2.5毫升缓冲液或血液,气液O₂/CO₂平衡最多耗时不到13分钟。与标准薄膜测压计相比,在0 - 93千帕(0 - 700毫米汞柱)和2 - 20千帕(14 - 150毫米汞柱)范围内,pO₂和pCO₂的测量结果具有良好的一致性。注射器式测压计消除了样品在转移和运输至分析仪过程中的污染,从而使测压法对技术的依赖性降低。该装置简单易用,相对于现有的测压方法具有明显优势。

相似文献

3
Syringe injection pressure: a neglected factor in blood PO2 determination.
Ann Clin Biochem. 1990 Mar;27 ( Pt 2):147-51. doi: 10.1177/000456329002700211.
4
Type of solution and PCO2 measurement errors during tonometry.
Intensive Care Med. 1997 Jun;23(6):658-63. doi: 10.1007/s001340050390.
9
Monitoring gastric mucosal carbon dioxide pressure using gas tonometry: in vitro and in vivo validation studies.
Anesthesiology. 1997 Sep;87(3):504-10. doi: 10.1097/00000542-199709000-00008.
10
Effects of syringe type and storage temperature on results of blood gas analysis in arterial blood of horses.
J Vet Intern Med. 2007 May-Jun;21(3):476-81. doi: 10.1892/0891-6640(2007)21[476:eostas]2.0.co;2.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验